Water quality:
pH value: 7–9
Conductivity: 50–60 µS/cm
Total hardness: < 14 °dH
Filter: max 150 µm

Download PDF here

German instruction: page 52 – 55

English instruction:  page 110 – 113

French instruction:  page 172 – 175

Software Upgrade:

The software for the Pacemaker HF is memorized on a small processor board. For an update with new software, you need to change this board.
NOTE:

Before you start to change the board write down the voltage value you were using for your application.

 

Switch off the main switch of PowerCURE™

 

WARNING:
Wait 5 minutes, before you remove the cover. Capacitors with high voltage inside!

 

Remove the processor board carefully by hand. Do not use any tools for that.

 

Install the new processor board.

 

NOTE:
Be careful and make sure that you mount the new processor board correctly and that pins match the socket!! Watch from the side.

 

NOTE:
When you first start up, use a low voltage of approx. 80 V.
We hope the upgrade work was successful. Thank for your support!

Possible cause:

There might be a loose connection!

 

Check the connection of the plug J1 on the main board (see picture).

Check the voltage on this plug while running:

white -> black = 0.5 – 1.0 VDC

green – > black = 15 VDC

brown -> black = -15 VDC

 

If you don’t have 0.5-1.0V, please check the wiring of this cable, which goes to a small board inside the of the PM HF.

NOTE:
Check this plug only when the unit is switched off and discharged!